Figure 5-4 Power distribution - UL
Cable
CE UPS configuration
UL UPS configuration
Max
length
Recommended (1)
Type
Max gauge
Recommended (1)
Type
R rectifier
5 x 10 mm²
3ph+N+PE
AWG3
AWG6
3ph+N+PE 24 m (78ft
8in)
L out /load 5 x 10 mm²
3ph+N+PE
AWG3
AWG6
3ph+N+PE 24 m (78ft
8in)
(1) Size and type may be adapted locally per local regulation.
NOTE
Electrical contractor is responsible for providing the power cables from Hospital mains to
Large Display UPS and the ground cable from Hospital mains to Large Display Cabinet as
per local regulations.