continued
MIS
num-
ber
Cable
Assem-
bly
UL Style
Voltage
rating (V)
Max Volt-
age car-
ried (V)
Cable
diame-
ter
(mm)
Connector type
Bigger
Plug size
(mm)
28838A 5184554
Mac LAB Cable Standard Length
28761A 5179560
5.5 X-Ray ON Lamp Distribution
The fitter shall install an X-Ray On lamp in the Exam room, and in the control room if necessary, outside
the patient area, so that they are connected to the equipment through connection terminals.
NOTE
Innova systems with 21 & 31 cm detector provides 24 VAC for driving X-Ray ON line.
Hospital or contractor to provide 1.5 mm
2
(AWG 14) wires to connect to the system.
X-Ray ON lamp is active during X-Ray exposures.
WARNING
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO CONFORM TO
INTERVENTIONAL STANDARD IEC/EN 60601-2-43.
SIGNAL INDICATING THE X-RAY ON SHALL BE PERCEPTIBLE BY THE OPERATOR IN
ALL THE LOCATIONS DEFINED FOR THE PERSONNEL WHO MAY RECEIVE
SCATTERED RADIATION.
In the control room, an additional X-Ray on light must be installed if the console (VCIM) indicator cannot
be perceived by all the persons in the control room.
Electrical Requirements
172
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...