Table 2-13 D distance of
ANGIO / CARDIO
CARDIO / NEURO
Omega V Long
1278 mm (50.3 in)
1395 mm (54.9 in)
Omega V non motorized Long
1278 mm (50.3 in)
1395 mm (54.9 in)
NOTICE
Due to the plastic bushing used in the USA to protect cables from the sharp
edges of conduits it is necessary to place the cable conduit inside the table cable
access opening but the height of the outcoming conduit plus bushing is limited
to 12 mm (1/2 in).
NOTE
Refer to table Chemical anchors Pull out efforts and recommendations in
for pull out effort on each fixation bolts.
Water Pipe Requirements
Figure 2-48 Water Conduit location with “Above Grade” anchor kits
NOTICE
In some countries, depending on local regulations, it may be forbidden to run
electrical cables and water pipes in the same conduit. In this case, two separate
conduits are required.
Equipment Requirements
90
Innova
TM
IGS 6
5750182-1EN 3
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...