NOTE
(2) Special Humidity Instructions: The following parts can be can be shipped in standard
shipment conditions with the requirement that on arrival to installation site, and before
supplying power to these parts, they shall be kept in an environmental relative humidity
equal or lower than their specified capability, and that’s for a minimum of 48 hours.
• 3 kVA UPS 110 V & 220 V, 1 kVA UPS (Maximum specified relative humidity capability =
90%)
• LDM Monitor (Maximum specified relative humidity capability = 90%)
• CVI Injector (Maximum specified relative humidity capability = 85%)
• Diamentor (Maximum specified relative humidity capability = 80%)
• 1MP MX191 LCD Monitor (Maximum specified relative humidity capability = 80%)
Table 1-2 Atmospheric Pressure - Storage and transport
SUB-SYSTEM COMPONENTS
ATMOSPHERIC PRESSURE (hPa)
MIN
MAX
Frontal Positioner
525
1013
Lateral Positioner
525
1013
Omega Table
48
1220
TSUI (TSSC, Smart box, Smart handle, TPD)
500
1010
Footswitch
500
1060
Monitor LCD Eizo
540
1053
Monitor LDM
200
1050
Cabinets C1 Frt, C1 Lat, C2
500
1060
UPS 3 kVA, 1 kVA
697
1130
Power Distribution Box (PDB) CE
500
1060
Power Distribution Box (PDB) UL
500
1060
UL Fluoro UPS cabinet (*)
500
1013
CE Fluoro UPS cabinet (*)
500
1013
LDM cabinet LD core
700
1060
Detector conditioner
120
1060
Tube chiller Coolix 4100
500
1060
VCIM
525
1013
General Requirements
5750182-1EN 3
Innova
TM
IGS 6
27
Summary of Contents for Innova IGS 6
Page 9: ...Page intentionally left blank...
Page 57: ...Figure 2 11 Omega V Table dimensions Equipment Requirements 46 InnovaTM IGS 6 5750182 1EN 3...
Page 61: ...Figure 2 14 Gas box outlets Omega Table Equipment Requirements 50 InnovaTM IGS 6 5750182 1EN 3...
Page 69: ...Figure 2 22 C2 Cabinet dimensions Equipment Requirements 58 InnovaTM IGS 6 5750182 1EN 3...
Page 147: ...Page intentionally left blank Environmental Requirements 136 InnovaTM IGS 6 5750182 1EN 3...
Page 158: ...Figure 5 6 PDB Schematic CE 1 2 Electrical Requirements 5750182 1EN 3 InnovaTM IGS 6 147...
Page 159: ...Figure 5 7 PDB Schematic CE 2 2 Electrical Requirements 148 InnovaTM IGS 6 5750182 1EN 3...
Page 161: ...Figure 5 9 PDB Schematic UL 1 2 Electrical Requirements 150 InnovaTM IGS 6 5750182 1EN 3...
Page 194: ...InnovaTM IGS 6...