SECTION 35
EVT PRO 2.0
®
ESTABLISHING THE COMMUNICATIONS DATA LINK
SECTION 36
EVT PRO 2.0
®
SOFTWARE UPDATES
Upon start-up of the EVT PRO 2.0
®
Hand Pump Case electronics and the EVT PRO 2.0
®
laptop, the two devices should automatically establish the Wi-Fi communications data link.
This automatic link-up can take up to 1 minute. However, sometimes it may be necessary
to manually establish communications. Follow the procedure below in order to manually
establish the Bluetooth communications data link:
1. Make sure the Hand Pump Case electronics is switched on.
2. After switching on the Hand Pump Case electronics, the Main Power Button Wi-
Fi Status Light (Refer to Figure 12.1 for location) will start flashing red.
3. Manually
reset
the
Hand
Pump
Electronics
by
pressing
the
Wi-Fi
Communications Linkup Button for 5 seconds.
4. The Electronics has now been reset, and the red status light will flash red.
5. When the Wi-Fi connection has been established, the status light will glow solid
green. This may take up to 1 minute.
Note
– If communications have not been established, repeat this procedure. If at that point
no connection has been established, please contact the Ventil Service Desk at
http://oms.ventil.nl/ClientLogIn.aspx
Upon start-up, if the EVT PRO 2.0
®
detects an internet connection, it will automatically
check for the latest version of software. If updates need to be made, then it will give the
user the option to install the downloaded new software version.
The software updating process is very user friendly and self explanatory. The system will
display the currently installed software version and the new downloaded software version.
The user can then decide if the new downloaded software version is to be installed by
following the on screen instructions.
SECTION 37
EVT PRO 2.0
®
CALIBRATION REQUIREMENTS
Any device which is used to measure and determine quantitative measurements for the
purpose of quality should be calibrated annually to certify results. This is typical language
used by any certifying body for ISO, PED, or other quality group such as NB or ASME.
As such it is recommended that the EVT PRO 2.0
®
be calibrated annually to ensure
compliance and adherence to ASME / NB and other standards globally.
EVT PRO 2.0
®
Operations and Technical Manual (Rev 1.0
– 2016)
NOT FOR COPY OR DISTRIBUTION
Page 47