SECTION 10
EVT PRO 2.0
®
HAND PUMP CASE
Figure 10.1
– Hand Pump Case
SECTION 11
EVT PRO 2.0
®
HAND PUMP CASE BATTERY
The EVT PRO 2.0
®
Hand Pump Case Battery is a new generation, rapid charging LiCAD
battery which slides into a receptacle on the side of the Hand Pump case, and is locked
into place using a retaining tab.
A stand alone charger is provided with universal AC
adapters for worldwide adaptability to the AC supply grid. Refer to Section 7.2 for parts
identification.
•
1 charge is good for 8 hours of continuous use
, or approx. 300 tests lasting
15 minutes each.
•
Operating temperature is between -4
˚F (-20˚C) and 140F (60˚C).
However,
the rule of thumb is to use environmental situation awareness.
Also,
temperature extremes will degrade the battery charge faster.
The EVT PRO 2.0
®
Hand Pump Case has several key advancements over the previous
version. The case itself is a soft sided case with internal frame. The case itself houses the
EVT PRO 2.0
®
upgraded pump assembly, electronics motherboard, battery supply, and
the Getac V110 laptop computer. Refer to figure 10.1.
The EVT PRO 2.0
®
Hand Pump Case with electronics is a state of the art assembly. It is
assembled from lightweight aircraft aluminum and consists of a hydraulic pump, a
directional control lever, and an electronics interface that sends test data to the computer
via a wireless Wi-Fi connection. The hand pump generates a maximum working pressure
of 1160 psi (80 bar).
The Hand Pump Case electronics can operate at temperatures between -
4˚F (-20˚C)
and 140F (60˚C). However, the rule of thumb is to use environmental situation
awareness. If a certain job site is too hot, cold or wet, for you to safely work, then
the job site is also too hot, cold, or wet for the EVT Pro 2.0
®
Hand Pump Case and
the electronics.
The Pump case has an average run time of 8 hours under battery power, or about 300
tests which are 15 minutes for each test.
The pump can be filled with any ISO46 or AW46 or higher rated hydraulic tool oil
.
This can be purchased locally from a lube shop
.
EVT PRO 2.0
®
Operations and Technical Manual (Rev 1.0
– 2016)
NOT FOR COPY OR DISTRIBUTION