[EN] English
12
K0467 Revision A
5.1 Connecting (Figure 7)
1. Before use, make sure the SELV power adaptor supplied with the
instrument is used (GE part number 191-370).
2. Install an accessible power isolator to use as the disconnecting device
in the power adaptor supply circuit.
3. The power adaptor input power supply range: 100 - 240VAC, 50 to
60Hz 660mA, Installation Category II.
Note:
The power adaptor must be supplied by a fused or overload-protected
power supply.
4. Connect the power adaptor to the instrument.
5. Switch the power supply on.
6. Check that the front panel display shows the power-up sequence (Ref:
K0470, User manual, section 3.2).
Note:
. After the power-up sequence, the instrument shows the default display on
the touch screen. The touch screen divides into a number of mimic keys.
5.2 Requirements for rack-mounted and panel-mounted instruments
1. Install an accessible power isolator to use as the disconnecting device
in the power adaptor supply circuit.
2. Set the power supply isolator to OFF.
3. Connect the power adaptor before sliding the instrument into the rack.
4. Set the power supply isolator to ON.
5. Check that the front panel display shows the power-up sequence (Ref: K0470, User man-
ual, section 3.2).
6 Communication Connections
Connect the applicable connectors into the rear panel communications ports and, if appropriate,
secure with the captive screws.
Note:
The RS232 and IEEE 488 interfaces are both enabled at power-up. Set the required
parameters in Supervisor Setup/communications
menu, (
(Ref: K0470, User manual, section
3.6).
Summary of Contents for Druck PACE1000
Page 38: ...K0467 R vision A 19 FR Fran ais...
Page 112: ...K0467 Revis o A 19 PT Portugu s...
Page 113: ...GE Digital Solutions K0467 A PACE General Electric Company 2010 GE General Electric Company GE...
Page 119: ...RU 6 K0467 A 2 4 3 1 2 NPT 2 3 3 3 1 1 25 MWP 3 2 210 3000 2 1 1 2...
Page 121: ...RU 8 K0467 A 3 6 210 3000 3 7 3 8 1 1 25 MWP 2 NPT...
Page 123: ...RU 10 K0467 A 5 1 1 2 M3 x 6 2 M3 x 8 3 4 5 6 7 8 9 4 2 6...
Page 124: ...K0467 A 11 RU 6 1 1 2 3 1 4 5 6 2 5 GE 191 370...
Page 128: ...K0467 A 15 RU 6 5 8 1 2 3 4 IEEE 488 5 Supervisor Setup communications K0470 6 7 1 2 3...
Page 129: ...RU 16 K0467 A 8 IEEE 488 IEEE 488 1 6 7 8 9 15 1 5 9 2 654 71 5 1 4 1...
Page 130: ...K0467 A 17 RU 7 GE www gemeasurement com RMA RGA RMA PACE1000 GE 8 www gemeasurement com...
Page 131: ...GE Digital Solutions K0467 A PACE 2010 General Electric Company GE General Electric Company GE...
Page 137: ...ZH 6 K0467 A 2 4 3 1 2 UUT NPT 2 3 3 3 1 1 25 x MWP 3 2 210 bar 3000 psi 2 1 1 2...
Page 139: ...ZH 8 K0467 A 3 6 210 bar 3000 psi 3 7 3 8 1 1 25 x MWP 2 NPT...
Page 141: ...ZH 10 K0467 A 5 1 1 2 M3 x 6 2 M3 x 8 3 4 5 6 7 8 9 4 2 6...
Page 142: ...K0467 A 11 ZH 6 1 1 2 3 1 4 5 6 2 5 GE 191 370...
Page 146: ...K0467 A 15 ZH 6 5 8 1 2 3 4 IEEE 488 5 K0470 6 7 1 2 3...