Procedures (CH1): DC mV/Volts (source)
8.5
Do the procedure as follows:
1.
Connect the applicable calibration equipment (Ref: Table
8-1).
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration (Zero and +FS) for the applicable range:
mV (source) ranges
Volts (source) ranges
2000 mV
20 V
4.
Check the calibration is correct:
•
Select the applicable Millivolts or Voltage (measure)
function (ref Section 0). Apply the input values that are
applicable to the calibration:
mV: 0, 100, 200, 1000, 2000
Volts (V): 0, 5, 10, 15, 20.
•
Check the error is within limits (Ref: Table 8-6 or Table