Procedures (CH1): Resistance (source)
8.10
Do the procedure as follows:
1.
Connect the applicable calibration equipment (Ref: Table
8-1).
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration, for each range.
•
Range: 0-400 Ω
•
Range: 400Ω -2000Ω
•
Range: 2kΩ -4kΩ