Do the procedure as follows:
1.
Connect the applicable calibration equipment (Ref: Table
8-1).
2.
Set-up the equipment:
•
Function = TC (measure)
•
TC Type = K Type
•
CJ Compensation, Mode = Automatic
3.
Set the reference unit temperature: 0°C
4.
Let the equipment get to a stable temperature
(minimum: 1 hour since the last power on).
5.
Record the values that follow:
•
TC temperature given on the reference unit T (actual)
•
TC temperature given on the calibrator, T (measured)
•
CJ temperature given on the calibrator, CJ (measured)
6.
Calculate the CJ (Cal Value) as follows:
•
CJ (Cal Value) = CJ (measured) - T (actual) + T
(measured).
•
Use the calibration menu to do a one-point calibration
for the CJ (measure) function.
•
When the display shows “Sampling complete”, set the
correct Cal Value = CJ (Cal Value) above.