User Manual
Druck DPI620 Genii
4.
Check the calibration is correct:
•
Select the applicable AC mV or AC Volts (measure)
function (ref Section 0).
5.
Apply the input values that are applicable to the
Calibration.
•
AC mV: 10, 500, 1000, 2000
•
AC Volts: 5, 10, 20
6.
Check the error is within limits (Ref: Table 8-15 or Table
Table 8-15 AC mV (measure) error limits
Applied
AC mV
Calibrator
uncertainty
(mV)
Permitted
DPI 620 error
(mV)
10
0.12
2.5
500
0.2
3.1
1000
0.28
3.75
2000
0.44
5.0
Table 8-16 AC Volts (measure) error limits
Applied
AC Volts
Calibrator
uncertainty
Permitted
DPI 620 error
5
0.0018
0.03
10
0.0026
0.037
20
0.0042
0.050