Frequency calibration (source function)
Do the procedure as follows:
1.
Connect the applicable calibration equipment (Ref: Table
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Set-up the equipment as follows:
Frequency meter:
Gate time = one second
Druck DPI 620 Genii:
Waveform = Square;
Amplitude = 10 V
Frequency = 990 Hz
4.
Use the calibration menu (Ref: Section 8.1) to do a one-
point calibration.
5.
Check the calibration is correct.
06 June 2013
K0541 issue 1
Page 151 of 175