4.
Check the calibration is correct:
•
Select the applicable Current (source) function(Ref:
•
Apply the values that follow:
CH1/CH2:
0.2, 6, 12, 18, 24
•
Check the error is within limits (Ref: Table 8-3)
Table 8-3 Current (source) error limits
Source mA
Calibrator
uncertainty
(mA)
Permitted
DPI 620
error (mA)
±0.2
0.00008
0.0010
±6
0.00023
0.0016
±12
0.00044
0.0022
±18
0.0065
0.0028
±24
0.0012
0.0034