User Manual
Druck DPI620 Genii
7.
Check the calibration is correct
•
Select the TC (measure) function
8.
Check the equipment gives a TC temperature that is the
same as the temperature on the reference unit ±0.1°C
(0.2°F)
Procedures (CH1): AC mV/Volts (measure)
8.13
Do the procedure as follows:
1.
Connect the applicable calibration equipment (Ref: Table
8-1).
2.
Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3.
Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration for the applicable AC function.
•
Use the local power supply frequency.
•
For the AC mV (measure) function:
point 1 = 200.0 mVAC,
point 2 = 2000.0 mVAC
•
For the AC Volts (measure) function:
point 1 = 2.000 AC,
point 2 = 20.000 VAC
06 June 2013
K0541 issue 1
Page 165 of 175