User Manual
Druck DPI620 Genii
Table 8-6 millivolts (source) error limits
Source mV
Calibrator
uncertainty
(mV)
Permitted
DPI 620 error
(mV)
0
0.0001
0.008
100
0.00046
0.0125
200
0.0009
0.017
1000
0.003
0.1
2000
0.006
0.14
Table 8-7 Voltage (source) error limits
Source
V
Calibrator
uncertainty
(V)
Permitted
DPI 620 error
(V)
0
0.000004
0.00042
5
0.000019
0.0007
10
0.000034
0.00010
15
0.000049
0.00013
20
0.000064
0.0016
06 June 2013
K0541 issue 1
Page 149 of 175