System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
34
1-4
Cable Connections
1-4-1
Preliminaries
1. The following interconnection cable numbers refer to the MIS MAP and GROUNDING
INTERCONNECTING MAP included at the end of this manual.
2. The reference numbers between brackets ( ) refer to internal cables also described in the Service
Manuals.
1-4-2
Mis Chart
System MPH 50/65/80 with Compax 40E / SG 60 or SG100 / XT Suspension
Note:
Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:
−
Numbers on the left presented in grey squares are HLA numbers (CABLES #).
−
The
symbol is used to identify those connections which remain to be done during on site
installation.
−
The
symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.
DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS
MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.