System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
23
•
For components: refer to Inspection Report
2179651 TDR or 2167647 TDR or 2172166
TDR
1 Person
Bucky Voltage Settings:
110 or 220AC for SG60
−
100 Bucky
110 or 220AC for Compax 40E Bucky
220V for XT Suspension, RS85, C40E,
TLK, SG
ÍÍ
ÍÍ
A23
Step
Description
Done
Supporting Documents
ÍÍ
ÍÍ
Mark each box when done
From previous page
POWER ON AUTHORIZED
CALIBRATION OF SYSTEM COMPONENTS
Continues on next page
CONFIGURATION CHECK
*
IR
meets
expectations?
Reconfiguration is complete.
If more reconfiguration is necessary, proceed to step A24
No
Yes
•
For Bucky: refer to Inspection Report 2179651
TDR, section, 4.2 for 40E, section 6.3 for SG
VERY IMPORTANT:
Read Inspection Report (IR) ( Dir. 21722166TDR for Compax 40E/MPH/SG/RS 85; Dir.
2167647TDR for Compax 40E/MPH SG/XT; Dir. 2179651TDR for Compax 40E/MPH/SG/Tomo−Link) through
carefully before eventually proceeding with the MPH software configuration, Step A24 of this Steering Guide.
or to Inspection Report 2167647 TDR or
2172166 TDR, section 4.2 for 40E; section
5.3 for SG
Refer also to orange label fixed to the Bucky
WARNING
MAINLY FOR USE IN THE USA:
See IST 004
Summary of Contents for Compax 40E