System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
27
Step
Description
Done
Supporting Documents
ÍÍ
ÍÍ
From previous page
Continues on next page
ÍÍ
ÍÍ
1 Person
1 1/2 Hours
X
−
ray beam and Light beam alignment check
Center X
−
Ray beam on Table
Center X
−
Ray beam on Bucky Stand
CENTERING ADJUSTMENT
See C40E SM: Dir 2150295
−
100,
Chapter 3, section
SID DETENT ADJUSTMENT
Í
Í
(FINAL POSITIONING)
Horizontal SID Detent
1 Person
1/2 Hour
See XT SM Dir 46
−
019803, section 8
or Dir 2118910
−
100, section 12
or RS
−
85 SM Mounting Instructions
Dir. 2172805
−
100 Chapter 2.
and RS 85 Interface SM 2165897
−
100,
Chapter 1, section 5.
A31
A32
−
For XT see Job Card IST002
of this manual
−
For RS 85 see Job Card IST003 of this
manual
1 Person
ÍÍ
ÍÍ
(SA or SM) Collimator
lamp voltage
VERIFICATION
COLLIMATOR CALIBRATION
A33
See Ultranet SA + RAFL
−
S sm
Dir. Y00W20L.04 or Ultranet SM sm
Dir. Y00W1L.04 Chapter 4.
Refer to Inspection Report 2179651 TDR or
2167647 TDR, section 3.3
or
Inspection Report 2172166 TDR section 4.7
1/2 Hour
Horizontal SID Detent
Horizontal SID Indicators