Equipment setup
2001989-200E
ApexPro™
3-9
Install two new AA alkaline batteries when you begin monitoring a new patient.
Install two new AA alkaline batteries when the digital display starts flashing.
To install two new AA alkaline batteries:
1.
Locate the battery cover at the bottom back of the oximeter.
2.
Press the latch tab and lift up to open the battery compartment.
3.
Insert the batteries as indicated with the polarity signs within the battery
compartment.
4.
Close the battery cover.
Transmitter connection
WARNING
DUST COVERS — If the dust covers for the interface connectors
become detached from the transmitter, they may pose a choking
hazard for pediatric patients. Inspect the dust covers before each use
to verify that they are securely attached. If the dust covers become
detached and cannot be reinserted into their retaining slot, do not use
them on the transmitter, and keep them out of pediatric patients’
reach.
1.
Connect the non-sensor end of the SpO2 probe into the 9-pin connector on the
top of the oximeter.
2.
Plug one end of the interconnection cable into the
INTFC
connector on the
oximeter. Plug the other end into the
INTFC
connector (labeled
2
on its dust
cover) on the transmitter.
See Interconnection cables on page 3-10.
3.
Turn the oximeter on. The digital display turns on and the power LED (horizontal
bar) flashes. The digital display stays on for one minute.
Once connected, follow your unit’s protocol for attaching the transmitter and the
oximeter to the patient. A common method is to place them back-to-back in the same
pouch and belt them on the patient.
Dinamap PRO monitors
A DINALink serial cable is used to connect the transmitter to the Dinamap PRO 100–
400 series monitors. The interconnect cable connects to either of the interface ports
on the transmitter.
Summary of Contents for ApexPro
Page 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Page 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Page 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Page 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Page 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Page 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Page 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Page 86: ...5 18 ApexPro 2001989 200E Alarms ...
Page 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Page 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Page 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Page 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Page 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Page 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Page 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Page 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Page 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Page 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Page 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Page 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Page 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Page 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Page 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Page 217: ......