D-2
ApexPro™
2001989-200E
Troubleshooting
ECG
Arrhythmia troubleshooting
Problem: Inaccurate heart rate and/or false asystole
Solution:
Check ECG signal from patient:
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes.
Check amplitude of ECG waveform:
1.
Click on the patient’s
ECG
tab.
2.
Click on all ECG leads in the Display section and check for 0.5 mV amplitude at
normal (1X) size. (At least 0.5 mV amplitude is required for QRS detection.) For
borderline signals, validate on a graph.
3.
If amplitudes are low, electrodes may need to be repositioned or replaced.
Relearn arrhythmia:
1.
Click on the patient’s
ECG
tab.
2.
Click on the
Relearn
button.
IF PROBLEM CONTINUES: Change to
Single-Lead
ECG detection and processing:
1.
Click on the patient’s
ECG
tab.
2.
Click on
Single-Lead
in the
Lead Analysis
section.
3.
Click on the ECG leads in the Display section and change top ECG waveform to
display the lead with the greatest amplitude. (At least 0.5 mV amplitude is
required for QRS detection.)
Problem: False ventricular calls
Solution:
Check ECG signal from patient: (V leads may exhibit polarity changes
which may occasionally cause an inaccurate call.)
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes. (If V lead is a problem, move the V lead to another V
position.)
4.
Relearn ECG:
Click on the patient’s
ECG
tab.
Click on the
Relearn
button.
IF PROBLEM CONTINUES:
1.
Remove the V lead(s).
Summary of Contents for ApexPro
Page 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Page 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Page 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Page 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Page 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Page 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Page 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Page 86: ...5 18 ApexPro 2001989 200E Alarms ...
Page 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Page 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Page 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Page 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Page 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Page 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Page 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Page 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Page 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Page 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Page 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Page 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Page 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Page 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Page 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Page 217: ......