Section 3-6 - Completing the Set Up
3-6-3
Connection the I/O Rear Panel
Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (i.e., IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everyone who connects additional equipment to the signal input part or signal
output part of the LOGIQ S8 system configures a medical system, and is therefore responsible to ensure
that the system complies with the requirements of the valid version of IEC60601-1-1. If in doubt, consult
the technical service department or your local GE Healthcare representative.
Figure 3-6 External I/O Connectors - on Rear of System
Table 3-11 External I/O Connector - Description
Item
Connector Name
Section number
Description
USB Port
R1/R2: USB2.0 Full Speed (Isolated)
R3/R4: USB2.0 Full Speed (No isolated)
HDMI Connector
Connector for external Monitor.
Ethernet
LAN for InSite Connection (RJ45)
.
AUDIO
Connector for External Speaker.
Audio line out (3.5mm pin jack)
Summary of Contents for 5418099
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Page 417: ......