GE H
EALTHCARE
PROPRIETARY TO GE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
Section 3-10 - Printer Installation
3 - 53
3-10-1-2 UP-D897/D898 - Printer Settings (cont’d)
1.) Select
Portrait
for Orientation in
Utility
->
Connectivity
->
Service
and press
Save
in the bottom
left.
2.)
Select
Utility
->
System
->
Peripherals
.
3.)
Select the printer to adjust (UP-D897/D898) from the pull-down menu under Standard Printer
Properties. Click
Properties
.
4.) Select Properties from Printer pull-down menu.
5.) Click
Printing Preferences
at the bottom of Properties Window.
6.) Select Layout tab and select:
-
Paper:
1920x1280
-
Orientation:
Portrait
-
Interpolation Method:
Bilinear
7.) Select the Density Adjust tab and select:
-
Gamma:
TONE2
-
Sharpness =
0
; Dark =
0
; Light =
0
; Sharpness =
2
8.) For saving the adjusted printer settings click
Apply
and then
OK
.
9.)
Close the ‘Printers’ -window with the close button.
10.) Exit System Setup with
Save&Exit.
11.)Assign the Printer to the remote keys.
see:
Section 3-10-3 "Adding the printer to the system" on page 3-64