7 - 16
Section 7-3 - Gathering Troubleshooting Data (R1 to R4)
7-3-7 Capturing Network Logs with Network Sniffer (cont’d)
By IP and port: Select Capture filter.
4.) Select
New
.
5.) Edit name for example DICOM port 104.
6.) Enter the string with the following syntax: port xxx and host yy.yy.yy.yy, where xxx is the port
number of “My Computer” in the LOGIQ™ S8 and yy.yy.yy.yy is the IP address of the device under
test (PACS, server, etc.).
7.) Press
OK
.
Figure 7-15 Sniffer Capture Filter IP and Port
8.) Select
Start
.
9.) Minimize network sniffer window and initiate communication to the DICOM device (e.g. send
images to the storage device or query Worklist).