Section 7-4 - Service Desktop (CSD) - R1 to R4
7 - 21
7-4-3
Global Service User Interface (GSUI)
Internationalization
The user interface provided by the service platform is designed for GE personnel and is in English only.
There is no multi-lingual capability built into the Service Interface.
Service Login
1.) Local Access (R1 to R4) via Utility - Service.
2.) Service Login screen is displayed.
3.) As soon as the Common Service Desktop (CSD) is started, the Service
[Home]
Page appears
Access / Security
The service interface has different access and security user levels. Each user is only granted access to
the tools that are authorized for their use.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
Restart LOGIQ S8 after diagnostics
Always shutdown the system and reboot after a diagnostics session. A red message on the task bar will
remind you that the system needs to be restarted after a diagnostic run or disruptive mode is enabled,
before it is returned for customer use.
Table 7-1
Access Authorization
USER LEVEL
ACCESS AUTHORIZATION
PASSWORD
Operator
Authorized access to specified diagnostics, error logs and utilities.
Same acquisition diagnostic tests as GE Service.
uls
Administrator
uls
External Service
gogems
GE Service
Knowledge of the service level password.
rotating security
password
Summary of Contents for 5418099
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Page 417: ......