Section 3-13 - Setting up InSite Connection (R1 to R4)
3 - 85
3-13-6 Configuring Agent Configuration (cont’d)
6.) Configure the following mandatory fields as desired:
-
Continent
-
Country
-
City
-
State
-
Institution
The Agent Configuration Tool screen opens.
Figure 3-53 Completed Agent Configuration Tool screen
Must be PILOT or
PRODUCTION
Must show IP addresses
for the server selected
Recommended to
filter *.zip
Must be DISABLE, unless
a proxy server is provided
by the hospital
Must match
System ID