Section 4-3 - General procedures
4-3-8-1 Before moving the system (cont’d)
1.) Adjust the monitor and control panel using Up/Down control and Swivel control to their centered and
lowest positions. Refer to
4-3-6-1 To raise/lower the operator panel
operator panel from side to side
.
2.) Flip down the monitor and lock the monitor arm.
Figure 4-11 Flip down the monitor and lock the monitor arm
Figure 4-12 Moving/Transporting position
3.) Turn the system off, including the circuit breaker (see
Section 3-7 "Power On/Off" on page 3-17
and remove the plug from the wall.
4.) All cables from off-board peripheral devices must be disconnected from the console.
5.) Disconnect the footswitch from the console.
6.) Wind the power cable around the cable hook.
NOTE:
To prevent damage to the Power Cord,
DO NOT
pull excessively on the cord or make sharp
bends while wrapping.
7.) Connect all probes to be used while off site. Ensure that probe cables are out of the way from the
wheels and not protruding beyond the console. Use the probe management hooks located below
the Operator Panel to further secure the probe cables.
8.) Store all other probes in their original cases or in soft cloth or foam to prevent damage.
9.) Put the coupling gel in the gel holder.
10.)Ensure that no loose items are left on the console and unlock the wheels.
Summary of Contents for 5418099
Page 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Page 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Page 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Page 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Page 417: ......