13. Positioning the calibration model
A calibration model is required to calibrate the scanner. A calibration model is included in the scope of delivery.
Each calibration model is measured industrially and assigned these measurements.
If your calibration model is damaged and you are unsure whether this model can be used, GC will be pleased to advise
you and provide you with a replacement if necessary.
It is recommended to have the calibration model always at hand during operation.
1. Top
2. Bottom
3. Depressions
4. Measured values
5. Model number
1
2
3
4
5
INSERTING AND REMOVING THE CALIBRATION MODEL
The calibration model is to be handled in the same manner as a single jaw model. To fasten it, you require the object
holder with the knurled screw.
OBSERVE THE SAFETY INSTRUCTIONS!
•
If required, loosen the knurled screw to provide more room on the object holder.
•
Place the underside of the calibration model on the object holder.
•
Press the depressions of the calibration model slightly against the threaded pins of the object holder.
•
Tighten the knurled screw.
The calibration model is placed correctly when it is flush both at the front and the rear.
The impression with the measured values faces the floating stop.
•
Insert the object holder. When doing it, observe the software’s messages.
•
Remove the object holder as usual.