4.5.16. Format (Channel Two)
Activating the MENU switch fifteen times will display the default setting Format “F8N1” (8-N-1). Using
the UP/DOWN switches, the Format can be selected for the Modbus communication interface.
The selectable Formats are:
8-N-1, 8-E-1, 8-O-1, or 8-N-(bits- parity-stop bits).
Activating the ENTER switch by changing the status ON/OFF, will flash “SAVE” on the display. Confirm
the SAVE action by activating the ENTER switch once more. Activating the MENU switch while “SAVE”
is flashing, will discard the SAVE and switch the unit back to normal operation. If ENTER is activated,
the unit will display “DONE” to confirm that the changes have been saved and return to normal
operation. The operator can change the format settings and if desired move to the next item (Address)
in the menu structure by activating the MENU switch straight after the change. The change will be
saved at a later stage by activating the ENTER switch at any point in the menu structure. If no switch is
activated for 60 seconds, the detector will return to normal operation without saving the changed
settings.