Recording TMJ movement
51
Adjust the setscrews of the side arm in a way that the
axis needle solely rotates without drawing a translatory
movement on the flag.
3.5.6
Attaching the sensors
Only touch the flags on their borders. Avoid touching the black measuring surface to prevent it
from being soiled or scratched. Also make sure that recording mode is turned off while you
mount the flags and styli.
Insert the hexagon socket screws that are contained in
the extension set into the threads of the measuring flags.
Fixate the flags on the designated sockets of the upper
face bow. Avoid jamming anything between the side arms
and the flags in this process.