CADIAX compact
70
Possible cause of error
Solution
A measuring flag is soiled.
Due to the flag being soiled, there is no continuous
electrical contact between the tip of the stylus and
the measuring surface of the flag. Clean the flag
surface according to the provided cleaning instruc-
tions.
A stylus is outside the allowed measuring area of
the flag due to incorrect mounting.
Repeat mounting the face bow. Make sure that
both styli are positioned to stay on the black
measuring surface of the flag, also during medio-
trusive movements.
"Sliding chair effect" caused by surface friction
between stylus and flag.
The movement of the stylus across the flag surface
can cause a so called "sliding chair effect" to oc-
cur. During such, the tip of the stylus shortly lifts
up from the flag surface, causing electrical contact
to be lost. During application, this effect is often
accompanied by a squeaking sound during move-
ments.
To minimize this effect, apply a smallest possible
amount of medical petroleum jelly onto the measur-
ing surface of the flag. Make sure to clean the flag
after the recording.
A stylus is defective.
Contact an authorized GAMMA service center.
A flag is defective.
Contact an authorized GAMMA service center.
The CADIAX device is defective.
Contact an authorized GAMMA service center.
Recorded curves are too short on one side.
Possible cause of error
Solution
A measuring flag is soiled.
Due to the flag being soiled, there is no continuous
electrical contact between the tip of the stylus and
the measuring surface of the flag. Clean the flag
surface according to the provided cleaning instruc-
tions.
A stylus is outside the allowed measuring area of
the flag due to incorrect mounting.
Repeat mounting the face bow. Make sure that
both styli are positioned to stay on the black
measuring surface of the flag, also during medio-
trusive movements.
The right or left measuring sensors are not connec-
ted to the device.
Insert the plugs correctly.
A stylus is defective.
Contact an authorized GAMMA service center.
A flag is defective.
Contact an authorized GAMMA service center.
The CADIAX device is defective.
Contact an authorized GAMMA service center.