CADIAX compact
40
3.4.5
Attaching the sensors
Only touch the flags on their borders. Avoid touching the black measuring surface to prevent it
from being soiled or scratched. Also make sure that recording mode is turned off while you
mount the flags and styli.
Fixate the flags on the designated sockets of the upper
face bow. Avoid jamming anything between the side
arms and the flags in this process.
Fixate the styli with the respective screws in the retain-
ers of the lower face bow. The cable of the stylus should
hang downwards after it is fixated.
If the system is mounted properly and the lower jaw is in
centric position, the stylus tip should be located in the
area of the upper back quarter of the flag.
The styli should be positioned in a way that their axis protrude by about 1 cm to the outside.
This range of movement is, among other things, required for mediotrusive movements.
The circuit between stylus and flag should always remain closed, even when the system is dis-
mantled after use and the individual parts are stored in their packages. Improperly, jerkily break-
ing the circuit could lead to faults or disturbances in the system, e.g., broken cables.
Avoid touching the measuring surface of the flags and the stylus pins during the measurement
procedure and preparatory work. This also applies to the patient.