Table 6-1: Normal Values of Key Operating Parameters
Inspection Description
Should be:
1
On/Off Line Indicator (Analysis
1 Panel)
•
On Line (green)
2
Run/Stop Indicator
•
Run (green)
3
Relay Indicators (Analysis 1
Panel)
•
Status “Normal” (green)
•
Service “Normal” (green)
•
Mode “Run” (green)
•
Control “Auto” (green)
4
Fault Indicators (Indicators
Panel)
•
Cell Temperature (Green)
•
Probe Temperature (Green)
•
Measure Range (Green)
•
Reference Range (Green)
•
Dark Range (Green)
•
I/O Board (green)
5
Cell Temperature
(Typical Cell Temperature =
150
o
C/302
o
F) (Analysis 2
Panel)
•
Digital Indicator (within set point)
6
Probe Temperature
(Typical Cell Temperature =
150
o
C/302
o
F) (Analysis 2
Panel)
•
Digital Indicator (within set point). This
parameter is used only when a probe RTD is
installed and the Control with RTD function is
selected
7
Cabinet Temperature (Analysis
2 Panel)
•
Digital Indicator
8
Air Demand Output
•
Digital - Indicating -5% to + 5%
•
Graphic Showing trend, last value on right side
9
H
2
S and SO
2
•
Digital – Separate indications in mole %
•
Graphic Dual trace trend, last value on right
side
Summary of Contents for 943-TGX
Page 48: ...Revision 4 48 August 02 2019 Figure 4 13 Network Sub Panel...
Page 94: ...Revision 4 94 August 02 2019 Figure 9 3 Control Cabinet Detail...
Page 96: ...Revision 4 96 August 02 2019 Figure 9 5 AC Wiring Diagram...
Page 97: ...Revision 4 97 August 02 2019 Figure 9 6 DC Signal and Wiring Diagram...
Page 98: ...Revision 4 98 August 02 2019 This Page Intentionally Blank...