Revision 4
42
August 02, 2019
To edit a parameter, use the
Field Prev
or
Field Next
button to access the field and press
Enter. A dialog box similar to Figure 4-6 will be presented. Enter the desired value, select the
Okay button and press
Enter
.
When all desired changes have been made, select Save. Selecting Revert will keep the
parameter values that were used before the start of the editing process.
a)
Plant Factor
- This parameter specifies the
Plant Factor
term used in the analyzer’s
calculation of Air Demand. The
Plant Factor
is a gain term applied to the air demand
output used to adapt the process variable (Air Demand) to the plant’s front end air
adjustment capability.
Plant Factor
is defined as follows:
pf = -Q
tg
(100) / ([H
2
S]
ag
) (Q
ag
)
where:
Q
tg
= flow rate of the tail gas stream in moles/unit time (typical)
Q
ag
= flow rate of the acid gas feed stream in moles/unit time (typical)
[H
2
S]
ag
= molar concentration of H
2
S in the acid gas feed stream (typical)
PF
=
plant factor
The value for
Plant Factor
is typically in the order of -3.0.
b)
Operating Ratio
- Under stoichiometric concentrations of H
2
S and SO
2
([H
2
S] / [SO
2
] =
2/1) the Air Demand output becomes zero regardless of the plant specific gain factor (F)
when the operating ratio (R
op
) is two (2). When the plant specific gain factor (F) is
established for a particular process, the units of Air Demand become ‘percent change
required in process air’. (
i.e.,
a computed Air Demand of +1.5% means that the process
is 1.5% excess in air.) To achieve optimum performance of the Claus Sulfur Recovery
Plant, the Air Demand should be kept near zero.
c)
H2S Span
- This parameter specifies the span factor utilized for the H
2
S
concentration
display and output. This value should always be close to 1.0.
Any deviation of the H2S span parameter will impact the output calibration
of the analyzer.
d) SO
2
Span
-
This parameter specifies the span factor utilized for the SO
2
concentration
display and output. This value should always be close to 1.0.
Any deviation of the SO2 span parameter will impact the output calibration
of the analyzer
.
e)
COS Span
-
This parameter specifies the span factor utilized for the COS concentration
display and output. This parameter (when used) should always be close to 1.0.
f)
SVAP Span
- This parameter specifies the span factor utilized for the SVap
concentration. This parameter (when used) should always be close to 1.0.
g)
CS2 Span
- This parameter specifies the span factor utilized for the CS2 concentration.
This parameter (when used) should always be close to 1.0.
Summary of Contents for 943-TGX
Page 48: ...Revision 4 48 August 02 2019 Figure 4 13 Network Sub Panel...
Page 94: ...Revision 4 94 August 02 2019 Figure 9 3 Control Cabinet Detail...
Page 96: ...Revision 4 96 August 02 2019 Figure 9 5 AC Wiring Diagram...
Page 97: ...Revision 4 97 August 02 2019 Figure 9 6 DC Signal and Wiring Diagram...
Page 98: ...Revision 4 98 August 02 2019 This Page Intentionally Blank...