Separate buttons:
Note that the reset input is fail-safe and must therefore use a normally closed switch.
FCO750 CONNECTOR
START
RESET
START
RESET
I/P COMMON
+24V DC
If these signals are supplied from a PLC the Ready signal can be used to check that the start
and reset signals have been actioned, otherwise the minimum timings below should be
observed.
START
READY
Up to 30mS
20mS Min.
50mS Min.
RESET
Single Switch:
The start signal is permanently connected and the instrument is controlled by the reset input.
The reset input is fail-safe so the position shown is for the reset state. The switch needs to be
in the start position for the whole of the test. At the end of the test the switch may be returned
to the reset position. This method can not use the auto-reset or continue-on-start features.
FUNCTION
FCO770 CONNECTOR
START
RESET
I/P COMMON
START
RESET
+24V DC
FCO770 CONNECTOR