8-16
Chapter 8 Maintenance
Motor Replacement
5
Touch [Touch Panel Adjust] to switch to the next screen.
6
Touch the center part of "+" displayed on the screen.
When the next pointer is displayed, touch the pointer.
When all five (5) points are touched, the touch panel adjustment will automatically end.
Motor Replacement
When replacing the motor, it is necessary to remove the recorder block. After the replacement, make sure to adjust
the meshing of gear.
Tool: Phillips Screwdriver #2, E Ring Plier
1
Remove the recorder block.
Refer to "General Assembly Diagram" (
"General Assembly Diagram" P5-9).
2
Remove the motor.
Remove the motor from the recorder block.
Remove the E ring, gear, washer from the motor.
3
Adjust the meshing of gear.
bl5985
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...