8-4
Chapter 8 Maintenance
Self-Diagnosis (User Maintenance)
General Test 1
Test the LCD, touch panel, key/LED, sound, and status.
1
Select [General Test 1] on the "Maintenance" menu.
The "General Test 1" screen will be displayed.
2
Press the
(Start/Stop) key on the operation panel.
On the touch panel test screen, touch the cross-shaped marks and check that the marks change to circle.
The screen will switch each time the mark is touched. Check that there are no missing dots or uneven colors
in each screen.
3
Press the operation panel key and check that the corresponding key displayed on the screen is highlighted.
To end the key test, touch [Next].
4
Touch the buzzer mark, and check that the corresponding sound is generated.
bl5966
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...