8-8
Chapter 8 Maintenance
Self-Diagnosis (User Maintenance)
2
Selecting the printing speed and pattern will start the printing.
3
Check the printing speed, and check that there are no phase shift, missing dots, or uneven colors.
I/O Test
Test the I/O operation.
1
Select [I/O Test] on the "Maintenance" menu.
"I/O Test" menu will be displayed.
2
Connect the LAN cable to the LAN connector and connect to the PC using a HUB.
When "IP address" window is displayed by touching [LAN], enter the IP address of the electrocardiograph.
3
Touch [LAN] to start the test, and check that "OK" is displayed.
4
Set the USB memory to all the USB ports.
5
Touch [USB1], [USB2], [USB3] to start the test, and check that "OK" is displayed.
6
Touch [Back] > [Back] > [To exam] to return to the examination screen.
bl5973
6
3
5
Summary of Contents for CardiMax 2
Page 4: ...2 ...
Page 10: ...Contents ...
Page 26: ...Chapter 1 General Description Contents ...
Page 34: ...Chapter 2 Operational Description Contents ...
Page 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Page 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Page 48: ...Chapter 4 Spare Parts List Contents ...
Page 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Page 68: ...Chapter 6 Software Update Contents ...
Page 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Page 72: ...Chapter 7 Troubleshooting Contents ...
Page 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Page 82: ...Chapter 8 Maintenance Contents ...
Page 104: ...Chapter 9 Maintenance and Inspection Contents ...
Page 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Page 116: ...Chapter 10 Appendix Contents ...