8
-
9
C10-xx
Measured pulse amplitude is abnormal.
Cause
:
The cuff size does not match the patient.
Solution
:
Check if the cuff size matches the patient and the cuff is properly wrapped, and then
measure.
C11-xx
Starting inflation rapidly increases the pressure.
Cause 1
:
The cuff is pressed.
Solution 1
:
Check that the cuff is not pressed, and measure with the patient arms extended as
much as possible.
Solution 2
:
If error persists, air system may be clogged.
The “NIBP Unit Error (E****)” message is displayed on the main unit.
Cause
:
Errors occur for the following causes.
E01-1
:
Communication error with the NIBP unit
E01-2
:
RAM Test Error (Main CPU)
E01-3
:
ROM Test Error (Main CPU)
E08-01
:
Communication Error with Sub CPU
E08-02
:
Watchdog Timeout (Main CPU)
E08-03
:
Pressure Offset Error
E08:04
:
Pressure Comparison Failure
E08-05
:
Sub CPU Power Supply Failure
E08-06
:
Pressure Sensor 2 Power Supply Failure
E08-07
:
Pressure Sensor 1 A/D Reference Power Voltage Failure
E08-08
:
Rapid Exhaust Error
E08-09
:
Air hose identification error
E09-A
:
Maximum cuff pressure is exceeded.
E09-B
:
Inflation time has been exceeded.
E09-C
:
Quick mode timeout
E09-D
:
Measurement started during the long term mode pause period
E09-E
:
Measurement Time Over
E09-F
:
Timeout of pressure data transmission from main CPU
E09-G
:
Pressure Sensor 1 +5V Power Supply Failure
E09-H
:
Zero Calibration Timeout
E09-I
:
ROM Test Error (Sub CPU)
E09-J
:
RAM Test Error (Sub CPU)
E09-L
:
Clock transmission ceased.
E09-M
:
Communication Failure at Power ON
E09-N
:
Pressure Comparison Failure
E09-O
:
Maximum inflation time has been exceeded.
E09-Q
:
Measurement was started before zero calibration
E09-R
:
Zero Calibration Error
E09-S
:
Watchdog Timeout (Sub CPU)
E09-T
:
+5V Digital Power Supply Failure
E09-U
:
Main CPU Power Supply Failure
E09-V
:
Pump Control Signal Failure
E09-W
:
Quick Exhaust Valve Control Signal Failure
E09-X
:
Sub CPU Constant Exhaust Valve Control Signal Failure
E09-Y
:
Main CPU Constant Exhaust Valve Control Signal Failure
Solution
:
These errors can be cleared by pressing the Cancel System Error on the NIBP
setup menu. If the same message is displayed many times, a failure of the equipment
can be considered.
Summary of Contents for HS-8000 Series
Page 4: ...Blank Page ...
Page 20: ...xvi Blank Page ...
Page 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Page 24: ...1 4 Blank Page ...
Page 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Page 34: ...3 4 Blank Page ...
Page 62: ...4 28 Blank Page ...
Page 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Page 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Page 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Page 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Page 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Page 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Page 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Page 96: ...8 18 Blank Page ...