9
-
20
Replacement of NIBP Unit
●
Usable time of NIBP Unit
The NIBP unit can measure approximately 100,000 times.
Replace the NIBP unit to maintain proper function since the air may leak and the unit may not
correctly function if the unit is used for a long time.
Only as a guide, 100,000 measurements are possible.
●
Tool
・
Phillips screwdriver (Large)
●
Check the number of NIBP measurement
Displays the number of measurement since the beginning of use. Use only as a guide for NIBP unit
replacement.
1
Press the Menu
→
Maintenance
→
Usage Time keys.
The Usage Time window will be displayed.
The number of NIBP measurement will be
displayed.
●
Disassemble and assemble
Pay attention to wiring and tubing, and follow the procedure below to replace.
1
.
Remove the screws (x 3) on the rear side
and remove the “rear cover”.
2
.
Remove the screw (x1) on the top side,
disconnect the analog output connector,
and remove the "top panel” and then “R/L
side panel”.
Summary of Contents for HS-8000 Series
Page 4: ...Blank Page ...
Page 20: ...xvi Blank Page ...
Page 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Page 24: ...1 4 Blank Page ...
Page 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Page 34: ...3 4 Blank Page ...
Page 62: ...4 28 Blank Page ...
Page 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Page 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Page 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Page 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Page 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Page 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Page 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Page 96: ...8 18 Blank Page ...