No Check Item
Check Procedure
Criteria
12 Pulse
Sound
*NOTE
Using the comprehensive tester, input
ECG and check the Heartbeat rhythm
sound.
Correct heartbeat rhythm
sound can be confirmed.
13 Alarm
Function
*NOTE
Set the upper alarm setting to 120
Beats/Min and change base-line heart
rate signal from 80 Beats/Min to 160
Beats/Min. Set the lower alarm setting
to 60 Beats/Min and change base-line
heart rate signal from 80 Beats/Min to
30 Beats/Min. Measure the time form
the changed moment till when the
alarm is set off.
。
The time to set off the alarm
should be 12 seconds or less.
*NOTE
: Since functional performance depends on design or software, these items are not required items
in Maintenance Check.
Test if necessary.
No Check Item
Check Procedure
Criteria
5. Respiration
01 Respiration
display
sensitivity
Input 0.5Hz sinusoidal wave whose base
resistance changes 1.5k
Ω
/1
Ω
using the
comprehensive tester or reference
respiration signal generator and check the
displayed amplitude.
Check with I Lead and II Lead.
Displayed amplitude
should be within
10mm±2mm.
02 Respiration
Rate
display accuracy
Input the reference respiration signal to the
comprehensive tester or respiration
simulator and check the respiration rate
display.
Test at respiration reference load signal in
60 or 120 per minute.
*NOTE
Check with I
Lead and II Lead.
Measurement error
should be within ±5bpm
or less.
03 Frequency
Characteristic
*1
Input 0.5Hz, 1.5Hz, or 2.5Hz from the
comprehensive tester or reference
respiration signal generator and measure
frequency characteristic. Check with I Lead
and II Lead.
To the crest value when
0.5Hz is inputted, the
crest value should be
70% or more when the
frequencies are as
follows.
Adult mode: 1.5Hz
Neonate mode:
2.5Hz
*NOTE
: Since functional performance depends on design or software, these items are not required items
in Maintenance Check.
Test if necessary.
No Check Item
Check Procedure
Criteria
6. Temperature
01 Temperature
display error
Input temperature 20, 30, or 40
°
C to the
comprehensive tester and calculate the error
to displayed value.
Should be within ±0.2
°
C
between 25
°
C or more to
45
°
C or less.
Should be within ±0.4
°
C
between 0
°
C or more
and less than 25
°
C.
02 Temperature
Probe off detection
Check the display by connecting and
disconnecting the probe.
Displayed numeric
should disappear.
10
-
8
Summary of Contents for HS-8000 Series
Page 4: ...Blank Page ...
Page 20: ...xvi Blank Page ...
Page 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Page 24: ...1 4 Blank Page ...
Page 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Page 34: ...3 4 Blank Page ...
Page 62: ...4 28 Blank Page ...
Page 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Page 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Page 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Page 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Page 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Page 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Page 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Page 96: ...8 18 Blank Page ...