Analog Adjust
Adjusts ECG, RESP, and IBP functions.
Pressing the offset keys on each adjust screen adjusts the offset.
Pressing the Gain keys on each adjust screen adjusts the gain.
Lastly, pressing the Prev.Disp. key checks that the adjusted data is written and pressing the
YES key makes the new adjusted data effective. Pressing the NO key does not upgrade the
adjusted data.
Service tools/Measurement Device
For Analog Adjust, prepare the following adjusting jig and measurement device.
Item Model
Type Note
Comprehensive tester
MEDSIM300
FLUKE
ECG, RESP, BP
Standard voltage and
current generators
2554 or equivalent
Yokogawa
Electric
Corporation
ECG
The following adjusting jig can be used as the above used jig and measurement device. For how to
adjust when the following adjusting jig is used, refer to the Operation Manual of “DMJ patient
monitor adjusting device".
Item Model
Type Note
Patient Monitor Adjusting
Device
DMJ-4 or DMJ-1
Fukuda
Denshi
Co., Ltd.
ECG, RESP, BP
Summary of Contents for HS-8000 Series
Page 4: ...Blank Page ...
Page 20: ...xvi Blank Page ...
Page 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Page 24: ...1 4 Blank Page ...
Page 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Page 34: ...3 4 Blank Page ...
Page 62: ...4 28 Blank Page ...
Page 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Page 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Page 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Page 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Page 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Page 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Page 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Page 96: ...8 18 Blank Page ...