60
Chapter 6:
Replacement Procedures
Install Power in Aux Board PCBA
1
Connect the Docking to Power in Cable to the board being installed per
Figure 6.17 on page 59
2
Put the Power in Aux Board PCBA in place and secure with the Power in Nut. Tighten to 10.0 inch
pounds using the Power in Tool.
3
Install the Power in Cap and tighten to 10.0 inch pounds using the Power in Tool.
4
Install the ECG Flex Interconnect Cable on the Main PCBA and Power Button with ECG PCBA.
5
Install screw and tighten with a T8 bit to 5.5 inch pounds.
Main PCBA Replacement
Required Part
• P26956-XX Service Assembly, Main PCBA, Sonosite PX
Remove Main PCBA
Note: This procedure will also remove the ComE PCBA and Power Supply PCBA. Their removal will be a
part of this process.
For the following steps, reference
Figure 6.19 on page 61
1
Disconnect Left and Right Speaker connections.
2
Disconnect 10 Fan connections.
3
Remove the ECG Flex Interconnect Cable.
4
Disconnect the USB Aux Cable.
5
Disconnect the WiFi Dongle USB cable if present.
6
Disconnect the AB cable.
7
Disconnect the 5 SHIF Assembly Cables.
8
Remove the Docking to Main Board Cable.
9
Remove the SHIF to Main Board Power Cable.
10 Use a T20 bit to remove the 10 screws securing Main PCBA to Bottom Enclosure.
Note: Each screw also secures 3 items to the Main PCBA. Take care to retain all items.
• P24968, Spring Clip (9 each) The left front assembly will not contain this part
• P24967, Flat Washer (10 each)
• P25716, Upper Grommet (10 each)
See
Figure 6.18 on page 60
for visual of these items.
Figure 6.18 Screw Stack
Spring Clip
Flat Washer
Upper Grommet
Summary of Contents for Sonosite PX System
Page 1: ...i Sonosite PX Ultrasound System Service Manual ...
Page 2: ...ii ...
Page 4: ...iv ...
Page 14: ...6 Chapter 1 Introduction ...
Page 18: ...10 Chapter 2 Specifications ...
Page 36: ...28 Chapter 3 Safety ...
Page 46: ...38 Chapter 4 System Overview ...
Page 54: ...46 Chapter 5 Troubleshooting ...
Page 79: ...Chapter 6 Replacement Procedures 71 ...
Page 80: ...72 Chapter 6 Replacement Procedures ...
Page 82: ...74 Chapter 7 Maintenance ...
Page 90: ...82 Chapter 8 Performance Testing ...
Page 102: ...94 Appendix A Replacement Parts ...
Page 109: ...101 ...
Page 110: ...102 P25393 01 P25393 01 ...