54
Chapter 6:
Replacement Procedures
Figure 6.11 Display Hinge Tubes
Install Clinical Display
1
Install the left and right Hinge Tubes on new Clinical Display per
Figure 6.11 on page 54
.
2
Install the center Hinge Tube.
3
Carefully feed the Clinical Display cable through the Bottom Enclosure shown in
Figure 6.9 on page 53
.
4
Place the Clinical Display on top of Bottom Enclosure and line the hinges up with the screw holes per
Figure 6.10 on page 53
.
5
Install two screws in each hinge with a T20 bit and tighten to 5.5 inch pounds.
6
Connect the Clinical Display Cable to the Main PCBA, J3002 per
Figure 6.10 on page 53
.
7
Put Display Cable EMI Shield in place and secure the two screws with a T6 bit. Tighten to 2.0 inch
pounds.
8
Feed the Wireless Dongle USB Cable through the Rear Enclosure and snap Rear Enclosure in place
per
Figure 6.8 on page 52
.
9
Remove tape holding On/Off Button in place.
10 Connect the Wireless Dongle to the USB cable.
11 Install the Wireless Dongle onto the Rear Enclosure per
Figure 6.7 on page 51
.
12 Install the User Interface.
13 Install the Batteries.
Center Hinge Tube
Right Hinge Tube
Left Hinge Tube
Summary of Contents for Sonosite PX System
Page 1: ...i Sonosite PX Ultrasound System Service Manual ...
Page 2: ...ii ...
Page 4: ...iv ...
Page 14: ...6 Chapter 1 Introduction ...
Page 18: ...10 Chapter 2 Specifications ...
Page 36: ...28 Chapter 3 Safety ...
Page 46: ...38 Chapter 4 System Overview ...
Page 54: ...46 Chapter 5 Troubleshooting ...
Page 79: ...Chapter 6 Replacement Procedures 71 ...
Page 80: ...72 Chapter 6 Replacement Procedures ...
Page 82: ...74 Chapter 7 Maintenance ...
Page 90: ...82 Chapter 8 Performance Testing ...
Page 102: ...94 Appendix A Replacement Parts ...
Page 109: ...101 ...
Page 110: ...102 P25393 01 P25393 01 ...