Chapter 8:
Performance Testing
Pulsed Wave (PW) Doppler Imaging
Continuous Wave (CW) Doppler Imaging
Image Quality Verification Test/Livescan
• Products with replaced subassemblies, or products that have been otherwise disassembled, must
undergo an Image Quality Verification Test/Livescan.
• The Image Quality Verification Test/Livescan should be performed after successfully completing all
applicable performance tests listed prior in this chapter.
• The test is completed before returning the system to service.
• A certified sonographer must perform the test.
• The Livescan test performed is at the discretion of the Sonographer and will represent their acceptance
of a successful service event.
• Review all saved images and verify that the images are displayed properly.
Test PW
Doppler
Imaging
1
Attach the P5-1 transducer.
2
Press the
Doppler
button for the Doppler sample gate.
3
Press the
Doppler
button again for the Doppler spectral trace.
4
Place a large drop of ultrasound gel on the transducer lens.
5
Adjust the
Gain
control as necessary and then gently tap the top of the gel and
observe a reflection on the spectral trace and the sound from the speakers.
6
Press the
Freeze
button and then save the image. Press the
Freeze
button to return
to live imaging.
7
Press the
2D
key to return to 2D imaging.
Test CW
Doppler
Imaging
1
Attach the P5-1 transducer.
2
Press the
Transducer icon
on the user interface.
3
Select the
Cardiac
exam type.
4
Press the
Done
softkey.
5
Press the
Doppler
button for the Doppler sample gate.
6
Press the
PW
softkey to switch to CW Mode.
7
Press the
Doppler
key again for the Doppler spectral trace.
8
Place a large drop of ultrasound gel on the transducer lens.
9
Adjust the
Gain
control as necessary and then gently tap the top of the gel and
observe a reflection on the spectral trace and the sound from the speakers.
10 Press the
Freeze
buttonand then save the image. Press the
Freeze
button to return
to live imaging.
11 Press the
2D
key to return to 2D imaging.
Summary of Contents for Sonosite PX System
Page 1: ...i Sonosite PX Ultrasound System Service Manual ...
Page 2: ...ii ...
Page 4: ...iv ...
Page 14: ...6 Chapter 1 Introduction ...
Page 18: ...10 Chapter 2 Specifications ...
Page 36: ...28 Chapter 3 Safety ...
Page 46: ...38 Chapter 4 System Overview ...
Page 54: ...46 Chapter 5 Troubleshooting ...
Page 79: ...Chapter 6 Replacement Procedures 71 ...
Page 80: ...72 Chapter 6 Replacement Procedures ...
Page 82: ...74 Chapter 7 Maintenance ...
Page 90: ...82 Chapter 8 Performance Testing ...
Page 102: ...94 Appendix A Replacement Parts ...
Page 109: ...101 ...
Page 110: ...102 P25393 01 P25393 01 ...