Chapter 6:
Replacement Procedures
57
Install Speaker
1
Apply a small amount of RTV664 onto Speaker Housing per
Figure 6.14 on page 57
2
Feed Speaker Wire through Bottom Enclosure and plug into Main PCBA
3
Seat Speaker in Speaker Housing, pressing into the RTV applied in step 1
4
Remove adhesive backing from Speaker Gasket and apply per
Figure 6.14 on page 57
Figure 6.14 Speaker Installation
USB Aux Board Replacement
Required Part
• P24915-XX Service Assembly, USB AUX Board PCBA, Sonosite PX
Remove USB Aux Board
1
Disconnect USB Aux Board from Main PCBA per
Figure 6.15 on page 58
2
Use a T10 bit to remove the two screws securing the USB Aux Board.
3
Remove the USB Aux Board from the system.
Install USB AUX Board
1
Place USB Aux Board in the system
2
Use a T10 bit to install two screws per
Figure 6.15 on page 58
and tighten to 5.5 inch pounds.
3
Connect USB Aux Board to Main PCBA J3017.
Apply RTV
Speaker with Gasket