Chapter 6:
Replacement Procedures
65
Remove Dock PCBA
1
Remove the Main PCBA per “Main PCBA Replacement” on page 60.
2
Use a T30 bit to remove the 2 Dock Alignment pins per
Figure 6.25 on page 65
3
Use a T8 bit to remove the 7 screws.
4
The Dock PCBA is free to remove.
Figure 6.25 Dock PCBA
Install Dock PCBA
1
Put the Dock PCBA in place
2
Hand thread 2 Alignment Pins until screw head makes contact and is flush with Dock PCBA. If
necessary, wiggle the screws or board until they are flush.
Caution:Do not force Alignment Pins in place. Failure to align them correctly will lead to a damaged
board.
3
Use a T8 bit to install the 7 screws and tighten to 5.5 inch pounds.
4
Use a T30 bit to tighten the Alignment Pins to 20 inch pounds.
5
Install the Main PCBA.
6
Install the Clinical Display.
7
Install the User Interface.
SHIF PCBA Replacement
Required Part
• P25394-XX Service Assembly, SHIF PCBA
Remove SHIF PCBA
1
Remove the Main PCBA per “Main PCBA Replacement” on page 60.
7 Screws
Alignment Pins
Summary of Contents for Sonosite PX System
Page 1: ...i Sonosite PX Ultrasound System Service Manual ...
Page 2: ...ii ...
Page 4: ...iv ...
Page 14: ...6 Chapter 1 Introduction ...
Page 18: ...10 Chapter 2 Specifications ...
Page 36: ...28 Chapter 3 Safety ...
Page 46: ...38 Chapter 4 System Overview ...
Page 54: ...46 Chapter 5 Troubleshooting ...
Page 79: ...Chapter 6 Replacement Procedures 71 ...
Page 80: ...72 Chapter 6 Replacement Procedures ...
Page 82: ...74 Chapter 7 Maintenance ...
Page 90: ...82 Chapter 8 Performance Testing ...
Page 102: ...94 Appendix A Replacement Parts ...
Page 109: ...101 ...
Page 110: ...102 P25393 01 P25393 01 ...