background image

1-15

For Safe Operation

1

DR-ID1270 Operation Manual    897N120821E

Table 4

Recommended separation distances between

Portable and mobile RF communications equipment and the DR-ID1270

The DR-ID1270 is intended for use in the electromagnetic environment in which radiated RF disturbances are 

controlled.

The customer or the user of the DR-ID1270 can help prevent electromagnetic interference by maintaining a 

minimum distance between portable and mobile RF communications equipment (transmitters) and the DR-ID1270 as 

recommended below, according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d

 = 1.2

80 MHz to 800 MHz

d

 = 1.2

800 MHz to 2.5 GHz

d

 = 2.3

 0.01

 0.12

 0.12

 

0.23

 0.1

 

0.38

 

0.38

 

0.73

 1

 1.2

 1.2

 

2.3

 10

 

3.8

 

3.8

 

7.3

 100

 12

 12

 

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d in metres 

(m) can be estimated using the equation applicable to the frequency of the transmitter, where 

P

 is the maximum output 

power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1:   At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2:   These guidelines may not apply in all situations.

 

Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Summary of Contents for FDR ES DR-ID1270

Page 1: ...ed when operating it After reading this manual store it nearby the DR ID1270 so that you can see it whenever necessary DR ID1270 Operation Manual 6th Edition July 2018 For Safe Operation System Config...

Page 2: ...ii DR ID1270 Operation Manual 897N120821E...

Page 3: ...phy or angiography Note The above statements were determined by applicable medical device regulations which vary throughout the world These statements are subject to revision when additional clearance...

Page 4: ...ightning etc This system is classified as a medical device under EC Directive 93 42 EEC Process waste correctly as stipulated by local law or any regulations that apply Caution Rx Only in the United S...

Page 5: ...ditions 2 10 2 4 1 Definition of Patient Environment 2 10 2 4 2 Installation Precautions 2 11 2 4 3 Precautions for Installing the Access Point Optional 2 11 2 5 Wireless Specifications 2 12 Chapter 3...

Page 6: ...s Z 3 Z 2 1 Precautions for Exposure of a Subject in Relatively Large Contrast Z 3 Z 2 2 Precautions for Flat Panel Sensor Z 3 Z 2 3 Precautions for Assuring the Radiation Field Z 3 Z 2 4 Images Outpu...

Page 7: ...e electrical load exceeding the allowable limit WARNING Observe the following precautions when using the cables Do not touch the plug and connector with wet hands Otherwise electric shock may result c...

Page 8: ...oc lectrique risque de se produire WARNING Make sure to use the optional parts recommended by us Failure to use the optional parts recommended by us may result in damage to the equipment and or electr...

Page 9: ...ndommagement de l quipement suite une chute ou un impact Installation Precautions CAUTIONS Do not install the equipment in a location with the following conditions Where the temperature changes sharpl...

Page 10: ...generate electromagnetic noise and their cables CAUTIONS Do not install the power supply unit in a place where it may be contacted inadvertently In addition take care not to make contact with the pow...

Page 11: ...re complete system isolation never install any unauthorized items When it is necessary to install authorized optional items contact our official dealer or FUJIFILM Representative WARNING Keep the prod...

Page 12: ...fection etc for each patient to prevent infection Precautions for Charging the Battery Pack CAUTIONS Use the battery charger recommended by FUJIFILM Corporation For details on operations refer to the...

Page 13: ...njured Do not touch the terminal of the battery pack directly There is a risk of electric shock Do not connect the positive and negative terminals with a wire or any metal object Do not carry or store...

Page 14: ...to overheat emit smoke explode or ignite Do not put the battery pack in a microwave oven or high pressure container Otherwise the battery pack may be rapidly heated or damaged causing it to overheat e...

Page 15: ...ecessary to systematically perform maintenance and inspection For information on maintenance and inspection contact our official dealer or FUJIFILM Representative CAUTIONS Do not perform maintenance a...

Page 16: ...and power box According to the degree of protection against electrical shock Type B applied part According to the degree of protection against harmful ingress of water IPX0 The flat panel sensor confo...

Page 17: ...terference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by tuning the equipment off and on the user is encoura...

Page 18: ...r decreased immunity of the DR ID1270 5 The DR ID1270 should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the DR ID1270 should be observed to verify...

Page 19: ...ge IEC 61000 4 5 1kV line s to line s 2kV line s to earth 1kV line s to line s 2kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short...

Page 20: ...metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Page 21: ...output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 1 2 800 MHz to 2 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0...

Page 22: ...by a projection it may be damaged Do not apply strong pressure onto the flat panel sensor If applied the flat panel sensor deforms and the waterproof function may be compromised Do not pull the cable...

Page 23: ...the system Up to five flat panel sensors can be connected If you intend to use six or more flat panel sensors only the first five that were connected to the image processing unit can be used For this...

Page 24: ...e Before making an exposure make sure that exposure conditions most appropriate for this system are set Do not apply an excessive force to the exposure plane The sensor inside the flat panel sensor ma...

Page 25: ...Cleaning To clean the outer surfaces use a cleaning cloth tightly wrung out of commercially available ethanol CAUTIONS Be sure to turn off the power before cleaning each part of the device Do not use...

Page 26: ...panel sensor In case that the flat panel sensor is deformed make sure that X ray images are not adversely affected before continuing the use of the flat panel sensor The precautions below must also b...

Page 27: ...at does not produce heating effects causes no known adverse health effects Many studies of low level RF exposures have not found any biological effects Some studies have suggested that some biological...

Page 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...

Page 29: ...the power supply unit is used the system can be operated in conjunction with the X ray equipment and it can be used for exposures in wired mode When the Power supply unit is used up to two flat panel...

Page 30: ...DR ID1275SE 7 Since the external dimensions and thickness of the flat panel sensor are the same as those of existing cassette for general exposure compliant with ISO 4090 the flat panel sensor can be...

Page 31: ...d DR ID1275SE incorporate a CsI indirect panel Battery pack level indicator Name Description Status lamp ERROR POWER READY LINK Indicates the equipment status by LEDs READY Green On Exposure possible...

Page 32: ...er options refer to the respective operation manuals SE communication cable Name Description SE communication cable A cable used for connecting the flat panel sensor and the optional access point This...

Page 33: ...fication Label Exposure plane Flat panel sensor DR ID1272SE Serial Number Label DR ID1273SE Radio Law Certification Label Battery Cover Label DR ID1273SE Identification Label Exposure plane Flat panel...

Page 34: ...et for MP Identification Label DR ID 1200MP DR ID1270MP Caution Label 1 DR ID 1200MP DR ID1270MP Caution Label 2 For the types of connectable cables consult our official dealer or FUJIFILM Representat...

Page 35: ...z notre revendeur agr ou notre repr sentant FUJIFILM DR ID 1200PB DR ID1270PB Identification Label DR ID 1200PB DR ID1270PB Rating Label DR ID 1200PB DR ID1270PB Caution Label DR ID 1200PB Identificat...

Page 36: ...f with your household waste according to the WEEE Directive and your national law This product should be handed over to a designated collection point Improper handling of this type of waste could have...

Page 37: ...s au rebut avec les d chets m nagers conform ment la directive DEEE et la l gislation nationale en vigueur Ce produit doit tre remis un centre de collecte appropri Une manipulation incorrecte de ce ty...

Page 38: ...guration page 2 1 CAUTIONS In the X ray room do not install the power supply unit access point image processing unit and battery charger optional in areas where the user could easily trip over them Fa...

Page 39: ...communication may become unstable If this happens the messages below may appear These messages are only examples Example 1 Panel communication error Example 2 Connection to X ray unit has been disconn...

Page 40: ...C This value is FUJIFILM measuring result of wireless module and actual data rate may vary 7 Transfer Power 17 dBm or less 8 Modulation OFDM Provided by IEEE802 11n 9 Wireless Data Security Wireless D...

Page 41: ...interval IEEE802 11n 20 HT20 40 HT40 Transmission rate IEEE802 11n HT20 MCS0 15 HT40 MCS0 15 Modulation OFDM Output power Power Max 17 Frequency Tolerance 20 20 ppm MAX Reception sensitivity PER Pack...

Page 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...

Page 43: ...onnected Up to two flat panel sensors can be connected to one power supply unit in wired communication mode Up to one power supply units can be used 3 1 3 Connecting Disconnecting the Flat Panel Senso...

Page 44: ...the radiographic examination stand CAUTIONS Make sure that the flat panel sensor is installed in the radiographic examination stand securely CAUTIONS Be careful not to have your fingers caught when i...

Page 45: ...t the flat panel sensor into the cassette receive with the green mark of the flat panel sensor up and then move it downwards Cassette receive Green mark 3 Set the flat panel sensor to the upper part o...

Page 46: ...h the tray back into place by using the handle after setting the flat panel sensor 4 Remove the flat panel sensor after use Hold the handle and pull out the tray Remove the flat panel sensor while pul...

Page 47: ...res for a maximum of approximately 500 images However the number varies depending on the usage conditions The capacity of the battery pack is displayed on the battery pack level indicator of the flat...

Page 48: ...ide marks misaligned may damage the connector terminal When attaching the battery pack make sure that the waterproof packing attached to the connector terminal of the flat panel sensor is aligned prop...

Page 49: ...n green Available time 60 minutes or more LED3 Blinking in green LED1 2 Lit in green Available time 30 minutes or more but less than 60 minutes LED2 Blinking in green LED1 Lit in green Available time...

Page 50: ...d connect the access point to the image processing unit CAUTIONS Use the optional access point by connecting it to the preset image processing unit and to the USB connector Do not use the optional acc...

Page 51: ...lting in an error the system startup Otherwise the system does not start up normally resulting in an error may be displayed in the connected devices status while information on radio wave strength is...

Page 52: ...CAUTIONS If the control cabinet is included in the system do not turn off the control cabinet with the main switch Shutdown operation may not be performed normally CAUTIONS When the system is shut do...

Page 53: ...egistered sleep mode is canceled automatically In addition when the SE cable is connected to the flat panel sensor or power box sleep mode is canceled automatically In sleep mode the operating time of...

Page 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...

Page 55: ...um cleaner or air duster clean each part with a slightly moistened soft cloth and then wipe off any moisture with a dry cloth See 2 2 Unit Names and the Functions page 2 3 CAUTIONS Be sure to turn off...

Page 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...

Page 57: ...on Atmospheric pressure 700hPa 1060hPa 2 Non operating Conditions Environmental conditions under which power can be supplied Temperature 5 C 35 C Humidity 10 RH 80 RH no dew condensation Atmospheric p...

Page 58: ...R ID1271SE DR ID1272SE DR ID1273SE DR ID1274SE DR ID1275SE 300kg 661 5 lb Local load DR ID1271SE DR ID1272SE DR ID1273SE DR ID1274SE DR ID1275SE 120kg 264 6 lb 40mm 1 6in Based on FUJIFILM measurement...

Page 59: ...8 0 1 Flat panel sensor DR ID1273SE 460 18 1 384 15 1 15 0 6 2 95 6 5 1 Flat panel sensor DR ID1274SE 460 18 1 460 18 1 15 0 6 3 65 8 0 1 Flat panel sensor DR ID1275SE 328 13 0 268 10 6 15 0 6 1 6 3 5...

Page 60: ...panel sensor DR ID1275SE The effective area of the at panel sensor is as shown in the gure below 20 0 8 20 0 8 288 11 3 230 4 9 1 17 45 0 7 20 45 0 8 Effective area DR ID1275SE Effective area 425 4 1...

Page 61: ...0821E Battery charger optional Width mm in Depth mm in Height mm in Weight kg lb Battery charger 92 5 3 6 259 10 2 56 2 2 0 6 1 3 AC adapter 153 6 0 67 2 6 35 1 4 Approx 0 51 1 1 153 6 0 67 2 6 35 1 4...

Page 62: ...nimum 2832x2836 pixels DR ID1272SE DR ID1274SE Image matrix size Minimum 1536x1920 pixels DR ID1275SE Gray scale Minimum 12 bit DICOM calibrated 5 Image Quality Evaluation Fujifilm typically conducts...

Page 63: ...SID to 100 cm 39 4 in or less and do not limit the radiation field to the bone region 1 only Alternatively set the SID according to the anatomical region and include the directly exposed area 2 More...

Page 64: ...flat panel sensor s battery pack may be reduced to half In addition the battery pack cannot be charged with an exposure menu registered even if the flat panel sensor has a wired connection For these r...

Page 65: ...ead characters or metals for measurement at the same position Z 2 3 Precautions for Assuring the Radiation Field CAUTIONS It is important to read the following before using the DR ID1270 digital detec...

Page 66: ...such moire effects sway the grid from side to side or use the Grid Pattern Removal Processing Software in conjunction with the grid with 40 lines Z 2 8 Precautions during Calibration Observe the follo...

Page 67: ...Access Point When making an exposure with the access point do not remove the USB cable until the image appears on the image processing unit and ShotReady lights green Also with the supplied Velcro tap...

Page 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...

Page 69: ...ions see O 3 Power Supply Unit page O 3 Power box A device supplying power to the flat panel sensor and connecting between the flat panel sensor and the control cabinet For the external view and preca...

Page 70: ...sensor to the SE cable or SE communication cable while it is stored in the SE storage case Do not store anything other than the flat panel sensor in the SE storage case Carefully carry the SE storage...

Page 71: ...the flat panel sensor and the inside of the power supply unit Power status LED Displays ON OFF of the power supply unit O 4 Power Box Power box Power status LED Be sure to connect the LAN cable 2m su...

Page 72: ...no license Do not use the Access Point near radios televisions mobile phones or PHS phones as far as possible If these devices are placed near the wireless LAN products audio or video noise may occur...

Page 73: ...terference to weather radar etc Even during start up the Access Point checks if there are any radar waves While the check is in progress communication with the Access Point is not available If a radar...

Page 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...

Page 75: ...l Every 1 years Number of exposures 90 000 It is recommended that the battery pack be replaced if the battery storage capacity becomes lower than 60 The battery pack should be replaced when the operab...

Page 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...

Reviews: