1.3.2 Before Exposure
The use of an air-conditioner may dramatically changes the temperature of the room where the
system is installed. This may cause dew condensation on the system, resulting in quality problems.
When an air-conditioner is used, change the temperature gradually to avoid temperature variation
in order not to cause dew condensation.
If an exposure is made with the front and rear of the flat panel
sensor facing the other way round, not only the re-exposure
is required but electric parts of inside the equipment may be
damaged.
Exposure plane of the
flat panel sensor
1.3.3 During Exposure
Before making an exposure, make sure that exposure conditions most appropriate for this system
are set.
Do not apply an excessive force to the exposure plane.
The sensor inside the flat panel sensor may be damaged,
and it may not be possible to make an exposure properly.
<Load restriction>
Entire surface load : DR-ID1271SE, DR-ID1272SE, DR-ID1273SE,
DR-ID1274SE and DR-ID1275SE :
300kg (661.5 lb)
Local load : DR-ID1271SE, DR-ID1272SE, DR-ID1273SE,
DR-ID1274SE and DR-ID1275SE :
120kg (264.6 lb) / ø40mm (1.6 in.)
(Based on FUJIFILM measurement specifications)
Use the flat panel sensor on a flat floor or platform.
When an excessive force is applied to the unit when it is tilted, the
sensor inside the flat panel sensor may be damaged.
Do not place a metal plate, etc., which blocks radio waves, before
the antenna. Otherwise, data may not be sent correctly from the flat
panel sensor.
Antenna
Antenna
Summary of Contents for FDR ES DR-ID1270
Page 2: ...ii DR ID1270 Operation Manual 897N120821E...
Page 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Page 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Page 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Page 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Page 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Page 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Page 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...