1-19
For Safe Operation
1
DR-ID1270 Operation Manual 897N120821E
When the SE communication cable is connected to the fl at panel sensor to make an exposure
on a bed, follow the precautions below. Otherwise a load may be applied locally to the SE
communication cable connectors, causing damage to the fl at panel sensor.
Make sure that the connector
does not protrude from the
edge of a bed
Do not place the connector
on a hard surface such as
the edge of a bed
Do not raise the flat
panel sensor by holding
only the connector
1.3.4 During Cleaning
To clean the outer surfaces, use a cleaning cloth tightly wrung out of commercially available ethanol.
CAUTIONS
● Be sure to turn off the power before cleaning each part of the device.
● Do not use an excessive amount of ethanol, as doing so may allow the liquid to enter from
the gap on the outer surfaces, resulting in the damage to the fl at panel sensor, or cause the
labels to come off.
● Do not use a solvent such as thinner or benzine, as it corrodes the outer surfaces.
● For other available disinfectants, consult our offi cial dealer.
● Use dehydrated ethanol or ethanol diluted with purifi ed water.
1.3.5 Storage
When the fl at panel sensor and the image processing unit are not in use, store them in a place
where they do not fall or drop.
Summary of Contents for FDR ES DR-ID1270
Page 2: ...ii DR ID1270 Operation Manual 897N120821E...
Page 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Page 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Page 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Page 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Page 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Page 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Page 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...