897N120551D
9-4
Chapter 9 Main Specification
9.2
Electromagnetic Compatibility (EMC) Information
This product complies with Class B of EN 55011 concerning radio-frequency emissions in the
combination of the 700 system scope.
In addition, in the combination of other endoscopes, this product complies with Class A. If this
product is used in domestic establishments with the Class A combination, electromagnetic
interference may occur in any devices. In such a case, the user is encouraged to try to correct
the interference according to “Chapter 1 Precautions.”
This product is intended for use in the electromagnetic environments specified below.
The customer or the user of this product should assure that it is used in such an environment.
Note
Combination with VP-7000
<Electromagnetic Emission Compliance Information and Guidance>
Emission standard
Compliance
Guidance
RF emissions
EN 55011
Group 1
This product uses RF (Radio Frequency)
energy for its internal function. Its RF
emissions are very low and are not likely to
cause any interference in nearby electric
equipment.
RF emissions
EN 55011
Class B
(Combination with
700 system scope
and VP-7000
Processor)
Class A
(Combinations
other than above)
[RF emissions]
With the Class B combination, this product
is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
With the Class A combination, this product
is intended for use in medical facilities and
commercial facilities.
Therefore, if this product is used in domestic
establishments, electromagnetic interference
may occur in any devices. In this case, it is
recommended to use this product according
to “Chapter 1 Precautions.”
Harmonic emissions
EN 61000-3-2
Class A
Voltage fluctuation/
flicker emissions
EN 61000-3-3
Applicable
BL7000_E-50_897N120551D.indb 4
2016/09/05 14:06:20
Summary of Contents for BL-7000
Page 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...
Page 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...
Page 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...
Page 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...
Page 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...